{"id":219,"date":"2022-10-20T16:32:20","date_gmt":"2022-10-20T16:32:20","guid":{"rendered":"https:\/\/www.science.smith.edu\/cfp\/?page_id=219"},"modified":"2022-10-31T20:49:23","modified_gmt":"2022-10-31T20:49:23","slug":"refurbishing-an-lcq-deca-xp-max","status":"publish","type":"page","link":"https:\/\/www.science.smith.edu\/cms\/resources\/instrument-repair-and-service\/refurbishing-an-lcq-deca-xp-max\/","title":{"rendered":"Refurbishing an LCQ Deca XP Max"},"content":{"rendered":"<p>The CMS has a Thermo Finnegan LCQ Deca XP Max used before the purchase of the Q Exactive HF-X in 2017. It is a linear ion trap instrument that can achieve only 4% the resolution possible on the QE HF-X and is significantly less sensitive. However the LCQ is a robust instrument with much lower gas consumption rates and is thus suitable as a general purpose and teaching instrument.<\/p>\n<p><img fetchpriority=\"high\" decoding=\"async\" class=\"alignnone size-medium wp-image-63\" src=\"https:\/\/www.science.smith.edu\/cfp\/wp-content\/uploads\/sites\/81\/2018\/08\/DSC_6576-300x200.jpg\" alt=\"\" width=\"300\" height=\"200\" srcset=\"https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2018\/08\/DSC_6576-300x200.jpg 300w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2018\/08\/DSC_6576-768x512.jpg 768w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2018\/08\/DSC_6576-1024x683.jpg 1024w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2018\/08\/DSC_6576-450x300.jpg 450w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2018\/08\/DSC_6576-272x182.jpg 272w\" sizes=\"(max-width: 300px) 100vw, 300px\" \/><\/p>\n<p>After a five year shutdown the instrument was cautiously restarted to identify components requiring service or replacement. Several issues were immediately identified. Cooling fans ran intermittently and were squeaking and clipping and the acquisition PC display did not work. This was traced back to a seized cooling fan and bulging capacitors in the graphics card. Fortunately removal of the graphics card allowed use of the integrated GPU. Instrument manager was not started by the PC and thus the tune page was unusable &nbsp;At this point the instrument was identified through the Ethernet data connection and data could be sent, however none was received.<\/p>\n<p>This first step was to restore communication with the instrument and a failure of the CMOS battery in the onboard instrument PC was suspected as this can result in BIOS errors during startup. Fortunately the LCQ onboard PC has VGA display and PS2 keyboard ports for diagnostics. To access these ports the instrument must be opened using by first loosening the large black bolt on the front of the instrument.<\/p>\n<p><img decoding=\"async\" class=\"alignnone wp-image-220 size-large\" src=\"https:\/\/www.science.smith.edu\/cfp\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221019_152740-1024x574.jpg\" alt=\"\" width=\"584\" height=\"327\" srcset=\"https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221019_152740-1024x574.jpg 1024w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221019_152740-300x168.jpg 300w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221019_152740-768x431.jpg 768w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221019_152740-1536x861.jpg 1536w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221019_152740-2048x1148.jpg 2048w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221019_152740-500x280.jpg 500w\" sizes=\"(max-width: 584px) 100vw, 584px\" \/><\/p>\n<p>With the instrument doors open the PC panel can be removed by unscrewing the panel cover on the right side. This is not required to connect the display and keyboard but is a good idea so the CPU fan can be checked as they have been known to seize in some instruments.<\/p>\n<p><img decoding=\"async\" class=\"alignleft wp-image-221 size-large\" src=\"https:\/\/www.science.smith.edu\/cfp\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221019_152723-574x1024.jpg\" alt=\"\" width=\"574\" height=\"1024\" srcset=\"https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221019_152723-574x1024.jpg 574w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221019_152723-168x300.jpg 168w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221019_152723-768x1370.jpg 768w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221019_152723-861x1536.jpg 861w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221019_152723-1148x2048.jpg 1148w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221019_152723-scaled.jpg 1436w\" sizes=\"(max-width: 574px) 100vw, 574px\" \/><\/p>\n<p>&nbsp;<\/p>\n<p>The keyboard and display ports are on the top of the panel. A PS2 mouse connector is in between them but is not usable.<\/p>\n<p><img loading=\"lazy\" decoding=\"async\" class=\"wp-image-246 size-large alignleft\" src=\"https:\/\/www.science.smith.edu\/cfp\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221018_163638-1-574x1024.jpg\" alt=\"\" width=\"574\" height=\"1024\" srcset=\"https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221018_163638-1-574x1024.jpg 574w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221018_163638-1-168x300.jpg 168w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221018_163638-1-768x1370.jpg 768w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221018_163638-1-861x1536.jpg 861w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221018_163638-1-1148x2048.jpg 1148w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221018_163638-1-scaled.jpg 1436w\" sizes=\"(max-width: 574px) 100vw, 574px\" \/><span style=\"font-weight: 300\">The PC booted but a CMOS battery failure error occurred. This following error was the reason for lack of communication with the instrument and can be bypassed by pressing F1 on the keyboard.<\/span><\/p>\n<p><img loading=\"lazy\" decoding=\"async\" class=\"alignnone wp-image-286 size-full\" src=\"https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221031_135557-scaled-e1667247586192.jpg\" alt=\"\" width=\"2092\" height=\"1401\" srcset=\"https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221031_135557-scaled-e1667247586192.jpg 2092w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221031_135557-scaled-e1667247586192-300x201.jpg 300w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221031_135557-scaled-e1667247586192-1024x686.jpg 1024w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221031_135557-scaled-e1667247586192-768x514.jpg 768w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221031_135557-scaled-e1667247586192-1536x1029.jpg 1536w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221031_135557-scaled-e1667247586192-2048x1372.jpg 2048w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221031_135557-scaled-e1667247586192-448x300.jpg 448w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221031_135557-scaled-e1667247586192-272x182.jpg 272w\" sizes=\"(max-width: 2092px) 100vw, 2092px\" \/><\/p>\n<p><span style=\"font-weight: 300\">The instrument booted successfully and communication was restored. Rebooting may need to be attempted several times with the analysis PC on or off until instrument manager starts automatically.<\/span><\/p>\n<p><img loading=\"lazy\" decoding=\"async\" class=\"alignnone wp-image-284 size-full\" src=\"https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221031_135629-scaled-e1667247705910.jpg\" alt=\"\" width=\"1939\" height=\"1433\" srcset=\"https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221031_135629-scaled-e1667247705910.jpg 1939w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221031_135629-scaled-e1667247705910-300x222.jpg 300w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221031_135629-scaled-e1667247705910-1024x757.jpg 1024w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221031_135629-scaled-e1667247705910-768x568.jpg 768w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221031_135629-scaled-e1667247705910-1536x1135.jpg 1536w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221031_135629-scaled-e1667247705910-406x300.jpg 406w\" sizes=\"(max-width: 1939px) 100vw, 1939px\" \/><\/p>\n<p><span style=\"font-weight: 300\"> The instrument was run for an hour to pump down before acquiring acceptable MS spectra with calibrant solution. The instrument was vented for to address the fan issues which we now knew were worth fixing as the instrument still functioned. The CMOS battery is a TL-5186 battery (the purple disk in the following image) and replacements are available, however it may be soldered to the board and the risk involved in changing it seems higher than keeping a keyboard connected to the motherboard for the rare times when the instrument is vented. The CPU fan spun freely but could be replaced as a precaution however this requires removing the CPU heat sink which was similarly avoided.<\/span><\/p>\n<p><img loading=\"lazy\" decoding=\"async\" class=\"alignnone wp-image-239 size-medium\" src=\"https:\/\/www.science.smith.edu\/cfp\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221018_163618-168x300.jpg\" alt=\"\" width=\"168\" height=\"300\" srcset=\"https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221018_163618-168x300.jpg 168w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221018_163618-574x1024.jpg 574w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221018_163618-768x1370.jpg 768w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221018_163618-861x1536.jpg 861w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221018_163618-1148x2048.jpg 1148w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221018_163618-scaled.jpg 1436w\" sizes=\"(max-width: 168px) 100vw, 168px\" \/><\/p>\n<p>There are five 120 mm cooling fans in the LCQ, two intake fans for the MS electronics, two exhaust fans for the turbomolecular pump and other vacuum chamber components, and one for the onboard PC. The onboard PC fan can be removed by unscrewing or unbolting the fittings on the bottom of the enclosure. These bolts are difficult to access and removing the black connector on the right side may be required to access one of the screw. Once the fan is free it can be disconnected by pulling out the black two pin plug on the fan body. The fans must be installed facing the same direction so it is best to keep notes.<\/p>\n<p><img loading=\"lazy\" decoding=\"async\" class=\"alignnone size-medium wp-image-242\" src=\"https:\/\/www.science.smith.edu\/cfp\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221018_164724-300x168.jpg\" alt=\"\" width=\"300\" height=\"168\" srcset=\"https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221018_164724-300x168.jpg 300w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221018_164724-1024x574.jpg 1024w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221018_164724-768x431.jpg 768w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221018_164724-1536x861.jpg 1536w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221018_164724-2048x1148.jpg 2048w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221018_164724-500x280.jpg 500w\" sizes=\"(max-width: 300px) 100vw, 300px\" \/><\/p>\n<p>Next the MS electronics and vacuum system fans can be removed by screwing the back housing plate<\/p>\n<p><img loading=\"lazy\" decoding=\"async\" class=\"alignnone size-medium wp-image-244\" src=\"https:\/\/www.science.smith.edu\/cfp\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221018_164910-300x168.jpg\" alt=\"\" width=\"300\" height=\"168\" srcset=\"https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221018_164910-300x168.jpg 300w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221018_164910-1024x574.jpg 1024w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221018_164910-768x431.jpg 768w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221018_164910-1536x861.jpg 1536w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221018_164910-2048x1148.jpg 2048w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221018_164910-500x280.jpg 500w\" sizes=\"(max-width: 300px) 100vw, 300px\" \/><\/p>\n<p>With these removed and the plate slid out the upper left side electronics cooling fans can be unscrewed.<\/p>\n<p><img loading=\"lazy\" decoding=\"async\" class=\"alignnone size-medium wp-image-238\" src=\"https:\/\/www.science.smith.edu\/cfp\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221018_162942-168x300.jpg\" alt=\"\" width=\"168\" height=\"300\" srcset=\"https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221018_162942-168x300.jpg 168w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221018_162942-574x1024.jpg 574w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221018_162942-768x1370.jpg 768w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221018_162942-861x1536.jpg 861w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221018_162942-1148x2048.jpg 1148w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221018_162942-scaled.jpg 1436w\" sizes=\"(max-width: 168px) 100vw, 168px\" \/><\/p>\n<p>Next the vacuum exhaust fans must be removed. This is more difficult as they are mounted on the inside of a removable chassis. First the catch screws (seen in the right side of the previous photo) on the bottom of this unit are removed. The chassis can then be slid out so the electronics connections can be removed. I disconnected the left and right BNC cables first.<\/p>\n<p><img loading=\"lazy\" decoding=\"async\" class=\"alignnone size-medium wp-image-233\" src=\"https:\/\/www.science.smith.edu\/cfp\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221018_155910-300x168.jpg\" alt=\"\" width=\"300\" height=\"168\" srcset=\"https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221018_155910-300x168.jpg 300w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221018_155910-1024x574.jpg 1024w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221018_155910-768x431.jpg 768w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221018_155910-1536x861.jpg 1536w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221018_155910-2048x1148.jpg 2048w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221018_155910-500x280.jpg 500w\" sizes=\"(max-width: 300px) 100vw, 300px\" \/><img loading=\"lazy\" decoding=\"async\" class=\"alignnone size-medium wp-image-234\" src=\"https:\/\/www.science.smith.edu\/cfp\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221018_155920-300x168.jpg\" alt=\"\" width=\"300\" height=\"168\" srcset=\"https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221018_155920-300x168.jpg 300w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221018_155920-1024x574.jpg 1024w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221018_155920-768x431.jpg 768w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221018_155920-1536x861.jpg 1536w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221018_155920-2048x1148.jpg 2048w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221018_155920-500x280.jpg 500w\" sizes=\"(max-width: 300px) 100vw, 300px\" \/><\/p>\n<p>Next the molex connectors to this PCB can be removed.<\/p>\n<p><img loading=\"lazy\" decoding=\"async\" class=\"alignnone size-medium wp-image-235\" src=\"https:\/\/www.science.smith.edu\/cfp\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221018_155933-300x168.jpg\" alt=\"\" width=\"300\" height=\"168\" srcset=\"https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221018_155933-300x168.jpg 300w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221018_155933-1024x574.jpg 1024w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221018_155933-768x431.jpg 768w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221018_155933-1536x861.jpg 1536w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221018_155933-2048x1148.jpg 2048w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221018_155933-500x280.jpg 500w\" sizes=\"(max-width: 300px) 100vw, 300px\" \/><\/p>\n<p>Finally the black plugs on the fan can be removed and the chassis removed for further work. The fans can easily be unscrewed at this point.<\/p>\n<p><img loading=\"lazy\" decoding=\"async\" class=\"alignnone size-medium wp-image-232\" src=\"https:\/\/www.science.smith.edu\/cfp\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221018_155450-300x168.jpg\" alt=\"\" width=\"300\" height=\"168\" srcset=\"https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221018_155450-300x168.jpg 300w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221018_155450-1024x574.jpg 1024w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221018_155450-768x431.jpg 768w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221018_155450-1536x861.jpg 1536w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221018_155450-2048x1148.jpg 2048w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221018_155450-500x280.jpg 500w\" sizes=\"(max-width: 300px) 100vw, 300px\" \/><\/p>\n<p>Now with all the fans removed they were spun to see which ones had bad bearings. The exhaust fans and PC cooling fans were partially or completely seized, leading to the sounds noted earlier. Rust and white mold was visible on the exterior of some fans. The exhaust and PC bearings may have gone bad first either due to condensation as they are in areas where air is cooling down (as evidenced by mold growth), or because the higher temperatures they are exposed to degraded their lubricant. These are ebmpapst 4394 fans and replacements are still available but are fairly expensive at ~$60 each and unfortunately the plug connectors are uncommon. Instead the fan bearings (twenty replacement 624ZZ bearings are available for under $20 on Amazon at the time of writing) were replaced as the motors seemed to be in good condition. I opted to change the bearings on all the fans (ten bearings total) instead of just those that had seized. <img loading=\"lazy\" decoding=\"async\" class=\"alignnone wp-image-222 size-large\" src=\"https:\/\/www.science.smith.edu\/cfp\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221017_130034-scaled-e1666282451544-1024x946.jpg\" alt=\"\" width=\"584\" height=\"540\" srcset=\"https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221017_130034-scaled-e1666282451544-1024x946.jpg 1024w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221017_130034-scaled-e1666282451544-300x277.jpg 300w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221017_130034-scaled-e1666282451544-768x710.jpg 768w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221017_130034-scaled-e1666282451544-325x300.jpg 325w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221017_130034-scaled-e1666282451544.jpg 1433w\" sizes=\"(max-width: 584px) 100vw, 584px\" \/><\/p>\n<p>After removing the stickers on the fans I large amounts of rust could be seen and the retaining clip on one fan had completely disintegrated.&nbsp;<\/p>\n<p><img loading=\"lazy\" decoding=\"async\" class=\"alignnone wp-image-223 size-large\" src=\"https:\/\/www.science.smith.edu\/cfp\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221017_130047-1024x574.jpg\" alt=\"\" width=\"584\" height=\"327\" srcset=\"https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221017_130047-1024x574.jpg 1024w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221017_130047-300x168.jpg 300w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221017_130047-768x431.jpg 768w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221017_130047-1536x861.jpg 1536w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221017_130047-2048x1148.jpg 2048w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221017_130047-500x280.jpg 500w\" sizes=\"(max-width: 584px) 100vw, 584px\" \/><\/p>\n<p>The retaining clips on the shaft can be removed with pliers and the blades slide out.<\/p>\n<p><img loading=\"lazy\" decoding=\"async\" class=\"alignnone size-medium wp-image-224\" src=\"https:\/\/www.science.smith.edu\/cfp\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221017_130126-300x168.jpg\" alt=\"\" width=\"300\" height=\"168\" srcset=\"https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221017_130126-300x168.jpg 300w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221017_130126-1024x574.jpg 1024w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221017_130126-768x431.jpg 768w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221017_130126-1536x861.jpg 1536w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221017_130126-2048x1148.jpg 2048w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221017_130126-500x280.jpg 500w\" sizes=\"(max-width: 300px) 100vw, 300px\" \/><\/p>\n<p>The bearings can be removed by inserting a punch or screwdriver through each side and pushing them out from the back. The fan blades and motor housing was cleaned with a damp cloth and dried to remove debris that might shorten the life of the replacement bearings. The new bearings can be gently inserted into the holes and a very small drop of lubricating oil added.<\/p>\n<p><img loading=\"lazy\" decoding=\"async\" class=\"alignnone size-medium wp-image-228\" src=\"https:\/\/www.science.smith.edu\/cfp\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221017_131733-168x300.jpg\" alt=\"\" width=\"168\" height=\"300\" srcset=\"https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221017_131733-168x300.jpg 168w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221017_131733-574x1024.jpg 574w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221017_131733-768x1370.jpg 768w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221017_131733-861x1536.jpg 861w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221017_131733-1148x2048.jpg 1148w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221017_131733-scaled.jpg 1436w\" sizes=\"(max-width: 168px) 100vw, 168px\" \/><\/p>\n<p>The fan blade and shaft can be slid back into the bearings. If the spring between the bearings and blades was removed during cleaning make sure to replace it first. Now new 4 mm retaining clips can be put on the shaft. I ordered a pack with many different sizes and a tool for around $20 however the latter was not needed for these small clips.<\/p>\n<p><img loading=\"lazy\" decoding=\"async\" class=\"alignnone wp-image-229 size-medium\" src=\"https:\/\/www.science.smith.edu\/cfp\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221017_132402-scaled-e1666283185965-300x287.jpg\" alt=\"\" width=\"300\" height=\"287\" srcset=\"https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221017_132402-scaled-e1666283185965-300x287.jpg 300w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221017_132402-scaled-e1666283185965-1024x978.jpg 1024w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221017_132402-scaled-e1666283185965-768x734.jpg 768w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221017_132402-scaled-e1666283185965-314x300.jpg 314w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221017_132402-scaled-e1666283185965.jpg 1433w\" sizes=\"(max-width: 300px) 100vw, 300px\" \/><img loading=\"lazy\" decoding=\"async\" class=\"alignnone wp-image-230 size-medium\" style=\"color: #333333;font-style: normal;font-weight: 300\" src=\"https:\/\/www.science.smith.edu\/cfp\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221017_133119-scaled-e1666283278423-207x300.jpg\" alt=\"\" width=\"207\" height=\"300\" srcset=\"https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221017_133119-scaled-e1666283278423-207x300.jpg 207w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221017_133119-scaled-e1666283278423.jpg 300w\" sizes=\"(max-width: 207px) 100vw, 207px\" \/><\/p>\n<p>The retaining rings can be gently pressed onto the shaft with fingers or pliers and rotated to ensure they properly seat in the groove.<\/p>\n<p><img loading=\"lazy\" decoding=\"async\" class=\"alignnone wp-image-231 size-full\" src=\"https:\/\/www.science.smith.edu\/cfp\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221017_133129-scaled-e1666283394709.jpg\" alt=\"\" width=\"246\" height=\"256\"><\/p>\n<p>Finally the fans were checked to ensure they spun freely and that there was minimal play in the rotor. The instrument was then reassembled and calibrant solution used to verify instrument functionality. The cost of refurbishing the fans and restoring the instrument to an operational state was under $50.<\/p>\n<p><img loading=\"lazy\" decoding=\"async\" class=\"alignnone wp-image-283 size-full\" src=\"https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221031_155934-scaled-e1667249327862.jpg\" alt=\"\" width=\"1433\" height=\"1747\" srcset=\"https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221031_155934-scaled-e1667249327862.jpg 1433w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221031_155934-scaled-e1667249327862-246x300.jpg 246w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221031_155934-scaled-e1667249327862-840x1024.jpg 840w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221031_155934-scaled-e1667249327862-768x936.jpg 768w, https:\/\/www.science.smith.edu\/cms\/wp-content\/uploads\/sites\/81\/2022\/10\/IMG_20221031_155934-scaled-e1667249327862-1260x1536.jpg 1260w\" sizes=\"(max-width: 1433px) 100vw, 1433px\" \/><\/p>\n","protected":false},"excerpt":{"rendered":"<p>The CMS has a Thermo Finnegan LCQ Deca XP Max used before the purchase of the Q Exactive HF-X in 2017. It is a linear ion trap instrument that can achieve only 4% the resolution possible on the QE HF-X &hellip; <a href=\"https:\/\/www.science.smith.edu\/cms\/resources\/instrument-repair-and-service\/refurbishing-an-lcq-deca-xp-max\/\">Continue reading <span class=\"meta-nav\">&rarr;<\/span><\/a><\/p>\n","protected":false},"author":262,"featured_media":0,"parent":218,"menu_order":0,"comment_status":"closed","ping_status":"closed","template":"","meta":{"_crdt_document":"","_monsterinsights_skip_tracking":false,"_monsterinsights_sitenote_active":false,"_monsterinsights_sitenote_note":"","_monsterinsights_sitenote_category":0,"ngg_post_thumbnail":0,"_themeisle_gutenberg_block_has_review":false,"footnotes":""},"class_list":["post-219","page","type-page","status-publish","hentry"],"yoast_head":"<!-- This site is optimized with the Yoast SEO plugin v27.3 - https:\/\/yoast.com\/product\/yoast-seo-wordpress\/ -->\n<title>Refurbishing an LCQ Deca XP Max - Center For Mass Spectrometry<\/title>\n<meta name=\"robots\" content=\"index, follow, max-snippet:-1, max-image-preview:large, max-video-preview:-1\" \/>\n<link rel=\"canonical\" href=\"https:\/\/www.science.smith.edu\/cms\/resources\/instrument-repair-and-service\/refurbishing-an-lcq-deca-xp-max\/\" \/>\n<meta property=\"og:locale\" content=\"en_US\" \/>\n<meta property=\"og:type\" content=\"article\" \/>\n<meta property=\"og:title\" content=\"Refurbishing an LCQ Deca XP Max - Center For Mass Spectrometry\" \/>\n<meta property=\"og:description\" content=\"The CMS has a Thermo Finnegan LCQ Deca XP Max used before the purchase of the Q Exactive HF-X in 2017. It is a linear ion trap instrument that can achieve only 4% the resolution possible on the QE HF-X &hellip; Continue reading &rarr;\" \/>\n<meta property=\"og:url\" content=\"https:\/\/www.science.smith.edu\/cms\/resources\/instrument-repair-and-service\/refurbishing-an-lcq-deca-xp-max\/\" \/>\n<meta property=\"og:site_name\" content=\"Center For Mass Spectrometry\" \/>\n<meta property=\"article:modified_time\" content=\"2022-10-31T20:49:23+00:00\" \/>\n<meta property=\"og:image\" content=\"https:\/\/www.science.smith.edu\/cfp\/wp-content\/uploads\/sites\/81\/2018\/08\/DSC_6576-300x200.jpg\" \/>\n<meta name=\"twitter:card\" content=\"summary_large_image\" \/>\n<meta name=\"twitter:label1\" content=\"Est. reading time\" \/>\n\t<meta name=\"twitter:data1\" content=\"10 minutes\" \/>\n<script type=\"application\/ld+json\" class=\"yoast-schema-graph\">{\"@context\":\"https:\\\/\\\/schema.org\",\"@graph\":[{\"@type\":\"WebPage\",\"@id\":\"https:\\\/\\\/www.science.smith.edu\\\/cms\\\/resources\\\/instrument-repair-and-service\\\/refurbishing-an-lcq-deca-xp-max\\\/\",\"url\":\"https:\\\/\\\/www.science.smith.edu\\\/cms\\\/resources\\\/instrument-repair-and-service\\\/refurbishing-an-lcq-deca-xp-max\\\/\",\"name\":\"Refurbishing an LCQ Deca XP Max - Center For Mass Spectrometry\",\"isPartOf\":{\"@id\":\"https:\\\/\\\/www.science.smith.edu\\\/cms\\\/#website\"},\"primaryImageOfPage\":{\"@id\":\"https:\\\/\\\/www.science.smith.edu\\\/cms\\\/resources\\\/instrument-repair-and-service\\\/refurbishing-an-lcq-deca-xp-max\\\/#primaryimage\"},\"image\":{\"@id\":\"https:\\\/\\\/www.science.smith.edu\\\/cms\\\/resources\\\/instrument-repair-and-service\\\/refurbishing-an-lcq-deca-xp-max\\\/#primaryimage\"},\"thumbnailUrl\":\"https:\\\/\\\/www.science.smith.edu\\\/cfp\\\/wp-content\\\/uploads\\\/sites\\\/81\\\/2018\\\/08\\\/DSC_6576-300x200.jpg\",\"datePublished\":\"2022-10-20T16:32:20+00:00\",\"dateModified\":\"2022-10-31T20:49:23+00:00\",\"breadcrumb\":{\"@id\":\"https:\\\/\\\/www.science.smith.edu\\\/cms\\\/resources\\\/instrument-repair-and-service\\\/refurbishing-an-lcq-deca-xp-max\\\/#breadcrumb\"},\"inLanguage\":\"en-US\",\"potentialAction\":[{\"@type\":\"ReadAction\",\"target\":[\"https:\\\/\\\/www.science.smith.edu\\\/cms\\\/resources\\\/instrument-repair-and-service\\\/refurbishing-an-lcq-deca-xp-max\\\/\"]}]},{\"@type\":\"ImageObject\",\"inLanguage\":\"en-US\",\"@id\":\"https:\\\/\\\/www.science.smith.edu\\\/cms\\\/resources\\\/instrument-repair-and-service\\\/refurbishing-an-lcq-deca-xp-max\\\/#primaryimage\",\"url\":\"https:\\\/\\\/www.science.smith.edu\\\/cfp\\\/wp-content\\\/uploads\\\/sites\\\/81\\\/2018\\\/08\\\/DSC_6576-300x200.jpg\",\"contentUrl\":\"https:\\\/\\\/www.science.smith.edu\\\/cfp\\\/wp-content\\\/uploads\\\/sites\\\/81\\\/2018\\\/08\\\/DSC_6576-300x200.jpg\"},{\"@type\":\"BreadcrumbList\",\"@id\":\"https:\\\/\\\/www.science.smith.edu\\\/cms\\\/resources\\\/instrument-repair-and-service\\\/refurbishing-an-lcq-deca-xp-max\\\/#breadcrumb\",\"itemListElement\":[{\"@type\":\"ListItem\",\"position\":1,\"name\":\"Home\",\"item\":\"https:\\\/\\\/www.science.smith.edu\\\/cms\\\/\"},{\"@type\":\"ListItem\",\"position\":2,\"name\":\"Resources\",\"item\":\"https:\\\/\\\/www.science.smith.edu\\\/cms\\\/resources\\\/\"},{\"@type\":\"ListItem\",\"position\":3,\"name\":\"Instrument Repair and Service\",\"item\":\"https:\\\/\\\/www.science.smith.edu\\\/cms\\\/resources\\\/instrument-repair-and-service\\\/\"},{\"@type\":\"ListItem\",\"position\":4,\"name\":\"Refurbishing an LCQ Deca XP Max\"}]},{\"@type\":\"WebSite\",\"@id\":\"https:\\\/\\\/www.science.smith.edu\\\/cms\\\/#website\",\"url\":\"https:\\\/\\\/www.science.smith.edu\\\/cms\\\/\",\"name\":\"Center For Mass Spectrometry\",\"description\":\"Mass Spectrometry and Chromatography at Smith College\",\"potentialAction\":[{\"@type\":\"SearchAction\",\"target\":{\"@type\":\"EntryPoint\",\"urlTemplate\":\"https:\\\/\\\/www.science.smith.edu\\\/cms\\\/?s={search_term_string}\"},\"query-input\":{\"@type\":\"PropertyValueSpecification\",\"valueRequired\":true,\"valueName\":\"search_term_string\"}}],\"inLanguage\":\"en-US\"}]}<\/script>\n<!-- \/ Yoast SEO plugin. -->","yoast_head_json":{"title":"Refurbishing an LCQ Deca XP Max - Center For Mass Spectrometry","robots":{"index":"index","follow":"follow","max-snippet":"max-snippet:-1","max-image-preview":"max-image-preview:large","max-video-preview":"max-video-preview:-1"},"canonical":"https:\/\/www.science.smith.edu\/cms\/resources\/instrument-repair-and-service\/refurbishing-an-lcq-deca-xp-max\/","og_locale":"en_US","og_type":"article","og_title":"Refurbishing an LCQ Deca XP Max - Center For Mass Spectrometry","og_description":"The CMS has a Thermo Finnegan LCQ Deca XP Max used before the purchase of the Q Exactive HF-X in 2017. It is a linear ion trap instrument that can achieve only 4% the resolution possible on the QE HF-X &hellip; Continue reading &rarr;","og_url":"https:\/\/www.science.smith.edu\/cms\/resources\/instrument-repair-and-service\/refurbishing-an-lcq-deca-xp-max\/","og_site_name":"Center For Mass Spectrometry","article_modified_time":"2022-10-31T20:49:23+00:00","og_image":[{"url":"https:\/\/www.science.smith.edu\/cfp\/wp-content\/uploads\/sites\/81\/2018\/08\/DSC_6576-300x200.jpg","type":"","width":"","height":""}],"twitter_card":"summary_large_image","twitter_misc":{"Est. reading time":"10 minutes"},"schema":{"@context":"https:\/\/schema.org","@graph":[{"@type":"WebPage","@id":"https:\/\/www.science.smith.edu\/cms\/resources\/instrument-repair-and-service\/refurbishing-an-lcq-deca-xp-max\/","url":"https:\/\/www.science.smith.edu\/cms\/resources\/instrument-repair-and-service\/refurbishing-an-lcq-deca-xp-max\/","name":"Refurbishing an LCQ Deca XP Max - Center For Mass Spectrometry","isPartOf":{"@id":"https:\/\/www.science.smith.edu\/cms\/#website"},"primaryImageOfPage":{"@id":"https:\/\/www.science.smith.edu\/cms\/resources\/instrument-repair-and-service\/refurbishing-an-lcq-deca-xp-max\/#primaryimage"},"image":{"@id":"https:\/\/www.science.smith.edu\/cms\/resources\/instrument-repair-and-service\/refurbishing-an-lcq-deca-xp-max\/#primaryimage"},"thumbnailUrl":"https:\/\/www.science.smith.edu\/cfp\/wp-content\/uploads\/sites\/81\/2018\/08\/DSC_6576-300x200.jpg","datePublished":"2022-10-20T16:32:20+00:00","dateModified":"2022-10-31T20:49:23+00:00","breadcrumb":{"@id":"https:\/\/www.science.smith.edu\/cms\/resources\/instrument-repair-and-service\/refurbishing-an-lcq-deca-xp-max\/#breadcrumb"},"inLanguage":"en-US","potentialAction":[{"@type":"ReadAction","target":["https:\/\/www.science.smith.edu\/cms\/resources\/instrument-repair-and-service\/refurbishing-an-lcq-deca-xp-max\/"]}]},{"@type":"ImageObject","inLanguage":"en-US","@id":"https:\/\/www.science.smith.edu\/cms\/resources\/instrument-repair-and-service\/refurbishing-an-lcq-deca-xp-max\/#primaryimage","url":"https:\/\/www.science.smith.edu\/cfp\/wp-content\/uploads\/sites\/81\/2018\/08\/DSC_6576-300x200.jpg","contentUrl":"https:\/\/www.science.smith.edu\/cfp\/wp-content\/uploads\/sites\/81\/2018\/08\/DSC_6576-300x200.jpg"},{"@type":"BreadcrumbList","@id":"https:\/\/www.science.smith.edu\/cms\/resources\/instrument-repair-and-service\/refurbishing-an-lcq-deca-xp-max\/#breadcrumb","itemListElement":[{"@type":"ListItem","position":1,"name":"Home","item":"https:\/\/www.science.smith.edu\/cms\/"},{"@type":"ListItem","position":2,"name":"Resources","item":"https:\/\/www.science.smith.edu\/cms\/resources\/"},{"@type":"ListItem","position":3,"name":"Instrument Repair and Service","item":"https:\/\/www.science.smith.edu\/cms\/resources\/instrument-repair-and-service\/"},{"@type":"ListItem","position":4,"name":"Refurbishing an LCQ Deca XP Max"}]},{"@type":"WebSite","@id":"https:\/\/www.science.smith.edu\/cms\/#website","url":"https:\/\/www.science.smith.edu\/cms\/","name":"Center For Mass Spectrometry","description":"Mass Spectrometry and Chromatography at Smith College","potentialAction":[{"@type":"SearchAction","target":{"@type":"EntryPoint","urlTemplate":"https:\/\/www.science.smith.edu\/cms\/?s={search_term_string}"},"query-input":{"@type":"PropertyValueSpecification","valueRequired":true,"valueName":"search_term_string"}}],"inLanguage":"en-US"}]}},"_links":{"self":[{"href":"https:\/\/www.science.smith.edu\/cms\/wp-json\/wp\/v2\/pages\/219","targetHints":{"allow":["GET"]}}],"collection":[{"href":"https:\/\/www.science.smith.edu\/cms\/wp-json\/wp\/v2\/pages"}],"about":[{"href":"https:\/\/www.science.smith.edu\/cms\/wp-json\/wp\/v2\/types\/page"}],"author":[{"embeddable":true,"href":"https:\/\/www.science.smith.edu\/cms\/wp-json\/wp\/v2\/users\/262"}],"replies":[{"embeddable":true,"href":"https:\/\/www.science.smith.edu\/cms\/wp-json\/wp\/v2\/comments?post=219"}],"version-history":[{"count":9,"href":"https:\/\/www.science.smith.edu\/cms\/wp-json\/wp\/v2\/pages\/219\/revisions"}],"predecessor-version":[{"id":287,"href":"https:\/\/www.science.smith.edu\/cms\/wp-json\/wp\/v2\/pages\/219\/revisions\/287"}],"up":[{"embeddable":true,"href":"https:\/\/www.science.smith.edu\/cms\/wp-json\/wp\/v2\/pages\/218"}],"wp:attachment":[{"href":"https:\/\/www.science.smith.edu\/cms\/wp-json\/wp\/v2\/media?parent=219"}],"curies":[{"name":"wp","href":"https:\/\/api.w.org\/{rel}","templated":true}]}}